Eye-Sync
More actions
| EYE-SYNC | |
|---|---|
| Basic Info | |
| VR/AR | Virtual Reality |
| Type | Head-mounted display |
| Developer | SyncThink, Inc. (renamed NeuroSync in 2022)[1] |
| Release Date | March 2016 (first generation)[2] |
| Website | https://neurosync.health/ |
| Versions | First generation (2016); second generation (June 2017); Pico Neo 3 Pro Eye version (2022, GUDID model X03)[3] |
| Requires | Companion tablet supplied with the system |
| Tracking | |
| Eye Tracking | Proprietary implementation of the "pupil-corneal reaction method" (FDA 510(k) summaries); 60 Hz camera speed in the Gear VR-based system (2020 510(k) summary); Tobii eye tracking at 90 Hz in the Pico Neo 3 Pro Eye version |
| Audio | |
| Camera | Two high-speed infrared eye cameras, lit by 12 LEDs centered at 850 nm (FDA 510(k) summaries) |
| Connectivity | |
| Connectivity | Bluetooth between headset and tablet; 4G cellular on the tablet (2020 510(k) summary) |
Eye-Sync (styled EYE-SYNC) is a head-mounted eye-tracking system that records and analyzes eye movements as an aid in assessing concussion. It was developed by SyncThink, Inc., a company founded in 2009 by neurosurgeon Jamshid Ghajar, which renamed itself NeuroSync in 2022.[2][1] The patient wears a virtual reality headset fitted with infrared eye cameras and follows a dot that moves in a circle; software on a companion tablet measures how closely the gaze stays synchronized with the target. The standard test takes less than a minute.[4][5]
The U.S. Food and Drug Administration (FDA) cleared the first version in January 2016 for recording, viewing and analyzing eye movements "in support of identifying visual tracking impairment". A second clearance in October 2021 added a narrower indication as an aid in the diagnosis of concussion within three days of a sport-related head injury in patients aged 17 to 24.[6][4] The hardware has moved across several headsets: an Oculus Rift DK2 used in 2016 demonstrations, a Samsung Gear VR modified by SensoMotoric Instruments, and from 2022 the Pico Neo 3 Pro Eye with built-in Tobii eye tracking.[7][8][9] Sports teams, universities, hospitals and the U.S. military have used it, and independent studies have reported mixed results on how well it detects concussion.[10][11]
Research background
SyncThink was founded in 2009 by Jamshid Ghajar, who was also president of the Brain Trauma Foundation, and the company developed the EYE-SYNC test in collaboration with that foundation.[2] In 2017 Ghajar was a clinical professor of neurosurgery at Stanford University and director of its Concussion and Brain Performance Center.[8][5] The test builds on a hypothesis that Ghajar and Richard Ivry published in 2008 in Neurorehabilitation and Neural Repair. They proposed that attention and memory problems after traumatic brain injury come from damage to white-matter connections between the prefrontal cortex, parietal lobe and cerebellum, which support the timing of anticipatory brain activity, so that the brain shifts "from a predictive to reactive mode".[12]
Brain Trauma Foundation researchers led by Jun Maruta then tested a circular visual tracking task. In a 2010 study of 17 people with chronic postconcussive syndrome and 9 controls, the variability of gaze error correlated with diffusion tensor imaging measures of white-matter tracts often damaged in mild traumatic brain injury, and with attention and working memory scores.[13] A 2013 paper with 47 healthy military volunteers set out the method: a target moving clockwise on a circle of 10 degrees radius at 0.4 Hz, recorded with a 500 Hz eye tracker, with metrics including the standard deviation of radial and tangential gaze error. Some participants made anticipatory saccades that put the gaze ahead of the target. Test-retest intraclass correlations ranged from 0.62 to 0.76.[14] A 2014 study found that 13 patients with acute mild traumatic brain injury tracked significantly worse than 127 controls, and that 26 hours of sleep deprivation also degraded performance in a different pattern.[15] These studies used laboratory eye trackers rather than the EYE-SYNC headset. A later paper published age- and sex-adjusted norms for the circular tracking task from 1,594 active-duty soldiers at Fort Hood.[16]
According to The Stanford Daily, the U.S. Department of Defense provided more than US$30 million to SyncThink in 2013 for clinical trials and development of EYE-SYNC.[8] SyncThink CTO Dan Beeler said in a 2016 interview that the military supported the foundational research, including a clinical study of about 10,000 people, half of them soldiers and half civilian athletes.[7]
Announcement and release
SyncThink announced the FDA clearance and commercial availability of EYE-SYNC on 22 March 2016, calling it the company's first medical device. In the release, Stanford's director of athletic training, Scott Anderson, said Stanford Sports Medicine was using the device as an investigational tool to screen athletes and make return-to-play decisions. The company also said it held 10 granted patents and a normative database of more than 10,000 people.[2] Stanford had been the field trial site in 2015, and its football program had used the system since the 2015-16 season.[7][5]
New Atlas described the system offered in March 2016 as a virtual reality-style headset hard-wired to a tablet.[17] At TechCrunch Disrupt in September 2016, SyncThink demonstrated the system on an Oculus Rift DK2 with SensoMotoric Instruments (SMI) eye trackers, connected to a Microsoft Surface Pro tablet; the whole system was packed in a briefcase and weighed about 15 pounds.[7] In an April 2016 Forbes contributor article, Ghajar said that "hundreds" of units had been sold and deployed in the first two months of sales. The goggles and software had cost about US$25,000 the previous year; Ghajar said the cost now varied with the number of users, from a few hundred to a few thousand dollars per month, as revenue came from monthly licensing fees for the software.[18] SMI had unveiled its eye-tracking version of the Gear VR, the SMI Mobile Eye Tracking HMD, on 22 March 2016.[19]
SyncThink released the second version on 26 June 2017. It added a vestibular test, in which the user moves the head while keeping the eyes on a fixed point, plus the SCAT5 sports concussion assessment tool and a cloud-connected analytics platform.[8][20] This version used a Gear VR headset with SMI eye-tracking hardware, since the stock Gear VR has no eye tracking.[8] First-generation customers received the new version along with training.[20]
Hardware
| Version | Headset | Eye tracking | Controller | Source |
|---|---|---|---|---|
| First generation (2016) | Head-mounted display (Oculus Rift DK2 in the 2016 demonstration) | Two high-speed infrared cameras and 12 LEDs centered at 850 nm; SMI trackers in the DK2 demonstration | Windows-based handheld (Surface Pro) | [6][7] |
| Second generation (June 2017) | Samsung Gear VR modified by SMI, with a phone as the display | Two high-speed infrared cameras, 12 LEDs at 850 nm; 60 Hz camera speed | Samsung 9.7-inch Android tablet with Verizon 4G | [8][4] |
| Pico version (2022) | Pico Neo 3 Pro Eye (Qualcomm Snapdragon XR2, standalone) | Tobii eye tracking at 90 Hz | Android tablet over Bluetooth | [9][3] |
The FDA summary for the 2016 clearance describes a head-mounted display containing the eye-tracking hardware and an integrated handheld that runs the test and stores results, with batteries for use away from a power source.[6] The 2020 submission describes the Gear VR-based system as an eye tracker "provided by SensoMotoric Instruments", a non-networked mobile device that fits inside the headset as the display, receives the eye data for post-processing, manages calibration, shows the binocular stimulus and talks to the tablet over Bluetooth, and a Samsung tablet that handles patient records, test control, analysis, reports and cloud synchronization. The system ships in a carry case with a user manual, a strap and cleaning and charging accessories, and the face mask has a foam cushion.[4] The Stanford Daily reported in July 2017 that SMI had recently been acquired by Apple.[8]
In March 2022 SyncThink said it would begin deploying the Pico Neo 3 Pro Eye version that month, with upgrade pathways for existing customers. SyncThink said the headset could start an assessment within ten seconds of power-up and that its spatial audio was used to automate test instructions.[9] The U.S. device database lists this version as model X03, made by NeuroSync Inc., with an onboard Snapdragon XR2 that processes the eye data and connects to the tablet over Bluetooth; the record, published in March 2025, lists it as in commercial distribution.[3] Another configuration used Pico Neo 2 headsets: in a recall posted by the FDA in July 2025, NeuroSync told customers to stop using EYE-SYNC units containing Pico Neo 2 headsets and to move to a Pico Neo 3 replacement system.[21]
How the test works
In the smooth pursuit test, the patient follows a target moving in a circle. The 2020 FDA summary specifies a 60-second recording of a target moving at 0.4 Hz on a 10-degree radius, in two sets of six cycles.[4] Ghajar described it as a 15-second test repeated twice, after which a report is produced.[5] The software compares gaze position with the target and reports two synchronization metrics, radial (spatial) variability and tangential (timing) variability, along with the mean position error of the eye.[22] Beeler said a common finding after head injury is a timing error, with the eyes jumping ahead to where the target will be.[7]
The 2020 submission lists four assessment paradigms on the headset: smooth pursuit, saccades, the vestibulo-ocular reflex (VOR) and VOR cancellation. The tablet app adds Vestibular/Ocular-Motor Screening (VOMS) self-report tools.[4] In 2017 SyncThink's website presented a screening algorithm that it said was adapted from Stanford Sports Medicine's concussion protocol. It paired EYE-SYNC eye tracking, EYE-SYNC vestibular testing and EYE-SYNC SCAT5 with a clinical cervical spine assessment and a migraine symptom assessment. The same page claimed that the test could identify signs of fatigue or poor sleep and could be used to monitor athletes' training workload.[22]
In late 2018 SyncThink's website described the package as Bluetooth virtual reality goggles, a smartphone and a tablet in a protective briefcase, with smooth pursuit, VOR and saccade assessments, tablet-based SCAT5, BESS and simple reaction time tests, HIPAA-compliant storage, unlimited tests with no per-test fees, and in-person and online training.[23] NeuroSync's current product page says each eye-tracking assessment takes less than 60 seconds and that results can be exported as PDF or through an API.[24]
Regulatory history
| Date | Event | Details |
|---|---|---|
| 29 January 2016 | 510(k) clearance K152915 | Class II nystagmograph (product code GWN, 21 CFR 882.1460); prescription use; cleared for "recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects". Predicate: Fall Prevention Technologies balanceback Mobile Intuitive VNG System. The summary states that no performance data were required.[6] |
| 31 July 2017 | FDA warning letter | FDA said SyncThink was promoting the device for cognitive and concussion assessment, a major change in intended use that needed a new clearance.[25] |
| February 2019 | Breakthrough Device designation | Granted for EYE-SYNC as an aid to concussion assessment.[26] |
| 2 October 2021 | 510(k) clearance K202927 | Class II traumatic brain injury eye movement assessment aid (product code QEA, 21 CFR 882.1455). Predicate: Oculogica EyeBOX (DEN170091).[4] |
| August 2022 | CE mark | Registered as a Class I medical device for the European Union and the United Kingdom, on the Pico Neo 3 Pro Eye version.[27][1] |
| July 2025 | Class 2 recall Z-2114-2025 | Posted by the FDA for 27 units of "NeuroSync EYE-SYNC containing Pico Neo 2 Headsets", distributed in New York, Texas and Georgia, for "Eye syncing issues causing the device to be inoperable". The recalling firm was NeuroSync, Inc. of Holliston, Massachusetts; the recall was listed as open.[21] |
The 2017 warning letter quoted marketing lines such as "In the clinic or at the point of injury: advanced cognitive assessment" and "Stanford Sports Medicine uses EYE-SYNC® technology currently to screen athletes for concussion", and told the company to stop distributing the device for those uses.[25] CEO Ernest Santin told MedCity News that the company had removed the references the FDA identified from its website and reviewed its other materials and videos against the clearance.[28]
The 2021 clearance covers use "as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI), within three days of sport-related head injury in patients 17-24 years of age in conjunction with a standard neurological assessment", by medical professionals qualified to interpret the results. A negative result corresponds to eye movements consistent with a lack of concussion; a positive result to eye movements that may be present in patients with concussion. The clinical data came from 1,069 competitive athletes aged 17 to 24, of whom 107 met the SCAT5 reference definition of concussion. The retrospective analysis reported 82% sensitivity, 93% specificity, a 56% positive predictive value and a 98% negative predictive value; assessments with poor eye-tracking quality, for which the device gives no result, were excluded.[4]
Use
Stanford's concussion protocol, which it submits to the NCAA each year, used EYE-SYNC as the first test for varsity athletes, and every Stanford athlete had a baseline test on record by 2017.[8] In August 2017 SyncThink agreed to supply the University of Texas sports medicine staff for all its athletics programs; Texas joined Stanford and Iowa State.[29] Indiana University was using it in a study on subconcussions.[5] By 2018 the company listed partnerships with the Pac-12 Conference, the University of Texas and the Golden State Warriors.[30] In 2019 NS Medical Devices listed Massachusetts General Hospital and the sidelines of Stanford, USC and Oregon State among its users, and quoted the company as saying it had worked extensively with NBA teams to track fatigue over the season.[31]
The Stanford Daily reported in 2017 that the U.S. Army had ordered several units and was testing them with troops in Afghanistan.[8] A U.S. Army Medical Research and Development Command team on non-invasive neuroassessment devices later identified EYE-SYNC, Oculogica's EyeBOX and Neuro Kinetics' IPAS as commercially available eye-tracking devices meeting its criteria toward being operationally effective in detecting traumatic brain injury in service members.[11]
Independent evaluations
- A Stanford study published in 2019 tested 150 healthy student athletes six times, before and after practice, with EYE-SYNC goggles. Intraclass correlations for the primary timing metric were 0.86 before practice and 0.88 after, and exercise changed scores little. The authors concluded that one measurement would be sufficient.[32]
- A prospective, case-controlled study of college athletes, published in 2022, compared sideline tests in 41 concussed athletes and 41 matched teammates without concussion. EYE-SYNC smooth pursuit tangential variability had an intraclass correlation of 0.70 and radial variability 0.47. For predicting a clinical concussion diagnosis, tangential variability reached 48% sensitivity and 58% specificity, with adjusted areas under the curve of 0.41 (tangential) and 0.47 (radial). Symptom scores performed best, and the King-Devick test was the best objective test.[10]
- A Walter Reed Army Institute of Research comparison of EYE-SYNC, EyeBOX and IPAS in adults aged 18 to 45 (63 recruited with acute mild traumatic brain injury, 34 with chronic injury and 119 controls) found classification performance ranging from acceptable to excellent in acute mild traumatic brain injury on the training data, with lower values on held-out validation data and less consistent results in chronic injury. The authors said device-specific differences in data quality, test length and ease of use had to be addressed before Department of Defense use.[11]
Company history
SyncThink was based in Boston at the time of the 2016 clearance.[2] In January 2018 it closed a US$3.5 million Series A-1 round led by Quadrant Management.[30] In January 2022 Magic Leap named SyncThink a participant in its Magic Leap 2 health Early Access Program, to work on clinical trials and explore the headset for vestibular disorders.[33] In March 2022 it expanded its partnership with Pico Interactive and said EYE-SYNC ran on three XR devices, with a fourth in development.[9]
On 20 September 2022 the company announced that it would rebrand as NeuroSync, with Gary Gregory as chairman and CEO and Ghajar as founder and CSO. The announcement claimed more than 30 peer-reviewed publications, 15 granted patents and over 4,000 providers.[1] As of October 2026 the product is still sold as EYE-SYNC on NeuroSync's website, which describes it as a virtual reality based platform for healthcare, biopharma, sport and military customers.[34]
See also
References
- ↑ 1.0 1.1 1.2 1.3 "SyncThink Rebrands as NeuroSync, Eyes Neurological Healthcare Market Growth". Healthcare Dive (press release). SyncThink. 2022-09-20. https://www.healthcaredive.com/press-release/20220920-syncthink-rebrands-as-neurosync-eyes-neurological-healthcare-market-growth. Retrieved 2026-10-21.
- ↑ 2.0 2.1 2.2 2.3 2.4 "EYE-SYNC Device That Assesses Abnormal Eye Movement, a Common Deficit After Concussion, Gains FDA Clearance". Marketwired (archived). SyncThink. 2016-03-22. https://web.archive.org/web/20161026173539/http://www.marketwired.com/press-release/eye-sync-device-that-assesses-abnormal-eye-movement-common-deficit-after-concussion-2108107.htm. Retrieved 2026-10-21.
- ↑ 3.0 3.1 3.2 "Device: EYE-SYNC (00860013612900)". AccessGUDID. U.S. National Library of Medicine. 2025-03-20. https://accessgudid.nlm.nih.gov/devices/00860013612900. Retrieved 2026-10-21.
- ↑ 4.0 4.1 4.2 4.3 4.4 4.5 4.6 4.7 "K202927: EYE-SYNC, 510(k) clearance letter and 510(k) summary". U.S. Food and Drug Administration. 2021-10-02. https://www.accessdata.fda.gov/cdrh_docs/pdf20/K202927.pdf. Retrieved 2026-10-21.
- ↑ 5.0 5.1 5.2 5.3 5.4 "Eye-Sync tracking technology helps diagnose concussions on the spot". CBS News. 2017-06-22. https://www.cbsnews.com/news/eye-sync-tracking-technology-diagnose-concussions-stanford/. Retrieved 2026-10-21.
- ↑ 6.0 6.1 6.2 6.3 "K152915: EYE-SYNC, 510(k) clearance letter and 510(k) summary". U.S. Food and Drug Administration. 2016-01-29. https://www.accessdata.fda.gov/cdrh_docs/pdf15/K152915.pdf. Retrieved 2026-10-21.
- ↑ 7.0 7.1 7.2 7.3 7.4 7.5 Kent Bye (2017-06-27). "#549: Using VR to Diagnose and Treat Concussions with SyncThink". Voices of VR Podcast. https://voicesofvr.com/using-vr-to-diagnose-treat-concussions-with-syncthink/. Retrieved 2026-10-21.
- ↑ 8.0 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 Deven Navani (2017-07-12). "SyncThink unveils latest version of EYE-SYNC". The Stanford Daily. https://stanforddaily.com/2017/07/12/syncthink-unveils-latest-version-of-eye-sync/. Retrieved 2026-10-21.
- ↑ 9.0 9.1 9.2 9.3 "SyncThink expands its partnership with Pico to integrate its EYE-SYNC technology with the Neo 3 Pro VR headset". Auganix. 2022-03-09. https://www.auganix.org/?p=11727. Retrieved 2026-10-21.
- ↑ 10.0 10.1 K. G. Harmon, B. M. Whelan, D. F. Aukerman, et al. (2022). "Diagnostic accuracy and reliability of sideline concussion evaluation: a prospective, case-controlled study in college athletes comparing newer tools and established tests". British Journal of Sports Medicine, vol. 56, no. 3, pp. 144-150. https://doi.org/10.1136/bjsports-2020-103840. Retrieved 2026-10-21.
- ↑ 11.0 11.1 11.2 J. King, C. Friend, D. Zhang, W. Carr (2024). "Comparative Performance of Three Eye-Tracking Devices in Detection of Mild Traumatic Brain Injury in Acute Versus Chronic Subject Populations". Military Medicine, vol. 189, suppl. 3, pp. 628-635. https://doi.org/10.1093/milmed/usae205. Retrieved 2026-10-21.
- ↑ J. Ghajar, R. B. Ivry (2008). "The predictive brain state: timing deficiency in traumatic brain injury?". Neurorehabilitation and Neural Repair, vol. 22, no. 3, pp. 217-227. https://doi.org/10.1177/1545968308315600. Retrieved 2026-10-21.
- ↑ J. Maruta, M. Suh, S. N. Niogi, P. Mukherjee, J. Ghajar (2010). "Visual tracking synchronization as a metric for concussion screening". Journal of Head Trauma Rehabilitation, vol. 25, no. 4, pp. 293-305. https://doi.org/10.1097/HTR.0b013e3181e67936. Retrieved 2026-10-21.
- ↑ Jun Maruta, Kristin J. Heaton, Elisabeth M. Kryskow, Alexis L. Maule, Jamshid Ghajar (2013). "Dynamic visuomotor synchronization: Quantification of predictive timing". Behavior Research Methods, vol. 45, no. 1, pp. 289-300. https://doi.org/10.3758/s13428-012-0248-3. Retrieved 2026-10-21.
- ↑ J. Maruta, K. J. Heaton, A. L. Maule, J. Ghajar (2014). "Predictive visual tracking: specificity in mild traumatic brain injury and sleep deprivation". Military Medicine, vol. 179, no. 6, pp. 619-625. https://doi.org/10.7205/MILMED-D-13-00420. Retrieved 2026-10-21.
- ↑ J. Maruta, L. A. Spielman, J. Ghajar (2023). "Visuomotor Synchronization: Military Normative Performance". Military Medicine, vol. 188, no. 3-4, pp. e484-e491. https://doi.org/10.1093/milmed/usab320. Retrieved 2026-10-21.
- ↑ Ben Coxworth (2016-03-22). "Eye movement monitor screens for concussion in 60 seconds". New Atlas (archived). https://web.archive.org/web/20251214101424/https://newatlas.com/eye-sync-concussion-test/42434/. Retrieved 2026-10-21.
- ↑ Eilene Zimmerman (2016-04-15). "Using Virtual Reality To Diagnose Concussions Before Sending Players Back Into The Game". Forbes (archived). https://web.archive.org/web/20250821140857/https://www.forbes.com/sites/eilenezimmerman/2016/04/15/using-virtual-reality-to-diagnose-concussions-before-sending-players-back-into-the-game/. Retrieved 2026-10-21.
- ↑ "SMI Unveils High Performance Eye Tracking on Samsung Gear VR". PR Newswire. SensoMotoric Instruments. 2016-03-22. https://prnewswire.co.uk/news-releases/smi-unveils-high-performance-eye-tracking-on-samsung-gear-vr-573080371.html. Retrieved 2026-10-21.
- ↑ 20.0 20.1 Ryan Mayer (2017-06-27). "SyncThink Introduces New, Upgraded Version Of Eye Sync Technology". CBS Local Sports (archived). https://web.archive.org/web/20171231063946/http://sports.cbslocal.com/2017/06/27/syncthink-introduces-new-upgraded-version-of-eye-sync-technology/. Retrieved 2026-10-21.
- ↑ 21.0 21.1 "Class 2 Device Recall NeuroSync EYE-SYNC (Z-2114-2025)". U.S. Food and Drug Administration, Medical Device Recalls database. 2025-07-11. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?id=214218. Retrieved 2026-10-21.
- ↑ 22.0 22.1 "EYE-SYNC". SyncThink (archived). 2017-09-17. https://web.archive.org/web/20170917025236/https://syncthink.com/eye-sync/. Retrieved 2026-10-21.
- ↑ "What's included". SyncThink (archived). 2018-11-23. https://web.archive.org/web/20181123052016/https://syncthink.com/included/. Retrieved 2026-10-21.
- ↑ "Product". NeuroSync. https://www.neurosync.health/product/. Retrieved 2026-10-21.
- ↑ 25.0 25.1 "SyncThink, Inc. - 498523 - 07/31/2017". U.S. Food and Drug Administration. 2017-07-31. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/syncthink-inc-498523-07312017. Retrieved 2026-10-21.
- ↑ Nick Paul Taylor (2019-02-20). "FDA grants breakthrough status to concussion-detecting VR headset". MedTech Dive. https://www.medtechdive.com/news/fda-grants-breakthrough-status-to-concussion-detecting-vr-headset/548756/. Retrieved 2026-10-21.
- ↑ Jürgen Thalmayer (2022-09-12). "SyncThink Secures CE Mark for Digital Health Platform". Wearable Technologies. https://wearable-technologies.com/news/syncthink-secures-ce-mark-for-digital-health-platform. Retrieved 2026-10-21.
- ↑ Stephanie Baum (2017-09-08). "FDA slaps SyncThink with warning letter over concussion detection claims". MedCity News. https://medcitynews.com/2017/09/fda-slaps-syncthink-warning-letter-concussion-detection-claims/. Retrieved 2026-10-21.
- ↑ Ryan Mayer (2017-08-30). "University Of Texas Adds EYE-SYNC Technology To Help Diagnose, Monitor Concussions". CBS Texas. https://www.cbsnews.com/texas/news/university-of-texas-adds-eye-sync-technology-to-help-diagnose-monitor-concussions/. Retrieved 2026-10-21.
- ↑ 30.0 30.1 Scott Hayden (2018-01-12). "SyncThink Closes $3.5 Million Series A Funding to Further Develop VR-based Concussion Detection". Road to VR. https://roadtovr.com/concussion-tech-closes-series-a-round/. Retrieved 2026-10-21.
- ↑ Anjuman Rahman (2019-07-31). "How SyncThink uses virtual reality to identify signs of concussion in athletes". NS Medical Devices. https://www.nsmedicaldevices.com/analysis/virtual-reality-brain-impairment/. Retrieved 2026-10-21.
- ↑ V. Sundaram, V. Y. Ding, M. Desai, A. Lumba-Brown, J. Little (2019). "Reliable sideline ocular-motor assessment following exercise in healthy student athletes". Journal of Science and Medicine in Sport, vol. 22, no. 12, pp. 1287-1291. https://doi.org/10.1016/j.jsams.2019.07.015. Retrieved 2026-10-21.
- ↑ "Magic Leap announces health technology participants in Early Access Program for its Magic Leap 2 enterprise AR platform". Auganix. 2022-01-13. https://www.auganix.org/?p=11363. Retrieved 2026-10-21.
- ↑ "NeuroSync". NeuroSync. https://neurosync.health/. Retrieved 2026-10-21.